Hydrom Instruction
ContactShopDeutsch
  • Introduction
    • Charging the Hydrom
    • Turn on the Hydrom
    • Start in Configuration-Mode
    • Access to the user interface
    • Connect to existing WiFi
    • Do PlainWaterCalibration
    • To set up the Hydrom as an MQTT publisher
    • Connect the Hydrom to Ubidots
    • Connect the Hydrom to Brewblox
    • Connect the Hydrom to Brewfather
    • Connect the Hydrom to Grainfather
    • Connect the Hydrom to Telegram
    • Send Measurements via Webhook
    • Connect the Hydrom to BierBot Bricks
    • Connect the Hydrom to Google Sheets
    • Connect the Hydrom to iRelays Display
    • Connect the Hydrom to Brewers Friend
  • Getting Started
    • Disinfect before use
    • Warnings (read before use)
    • Establish first connection to the hydrom
    • Indicator LEDs
  • Other settings
    • Changing the Hydrom name
    • Customize units
    • Factory Reset
    • Adjust Intervall
    • Activate headstand detection
    • Setting up the temperature compensation
  • Calibration
    • Check calibration status
    • Attitude Sensor Calibration
    • Calibration process
      • Manually change the formula
      • Reference method
  • Bluetooth Settings
    • IOS App
  • Troubleshooting
    • An Internet service does not receive any data
    • The Hydrom user interface cannot be accessed.
  • Firmware
Powered by GitBook
On this page
  • Step 3: Select Calibration
  • Step 4: Perform the measurements.
  • Session 1: 0° Brix/Plato solution
  • Session 2: 20° Brix/Plato solution
  • Session 3: 15° Brix/Plato solution
  • Session 4: 10° Brix/Plato Solution
  • Session 5: 7.5° Brix/Plato solution
  • Session 6: 5° Brix/Plato solution
  • Session 7: 2.5° Brix/Plato solution
  • Step 5: Confirm the results
  • The calibration is completed

Was this helpful?

  1. Calibration
  2. Calibration process

Reference method

Reference method is a calibration method that is considered to be very accurate. The density of the current liquid is measured and communicated to the Hydrom. The hydrometer then calculates the calibr

Precondition:

  1. User interface Access To make this setting you must have access to the user interface. If this is not the case, see here how to do it:Access to the user interface

  2. Equipment needed:

    1. Density verification meter

    2. Reference liquid (can be beer in the brewing process)

Step 3: Select Calibration

Step 4: Perform the measurements.

Session 1: 0° Brix/Plato solution

Create a 0° Brix solution by placing clear water (20°C) in a vessel.

Place the Hydrom in the vessel with the water.

When the instrument floats calmly, enter "0" in the Plato field and confirm by pressing "continue".

Session 2: 20° Brix/Plato solution

Create a 20° Brix solution by mixing 1816g of the solution from Pass 1 with 454g of sugar.

Place the Hydrom in the vessel containing the mixture.

When it is floating calmly and the sugar has completely dissolved then enter "20" in the Plato field and confirm by pressing "continue".

Session 3: 15° Brix/Plato solution

Create a 15° Brix solution by mixing 1668g of the solution from Pass 2 with 556g of water.

Place the hydrom in the vessel containing the mixture.

When it is floating calmly then enter "15" in the Plato field and confirm by pressing "continue".

Session 4: 10° Brix/Plato Solution

Create a 10° Brix solution by mixing 1453g of the solution from Pass 3 with 727g of water.

Place the hydrom in the vessel containing the mixture.

When it is floating calmly then enter "10" in the Plato field and confirm by pressing "continue".

Session 5: 7.5° Brix/Plato solution

Create a 7.5° Brix solution by mixing 1619g of the solution from Pass 4 with 540g of water.

Place the hydrom in the vessel containing the mixture.

When it is floating calmly then enter "7.5" in the Plato field and confirm by pressing "continue".

Session 6: 5° Brix/Plato solution

Create a 5° Brix solution by mixing 1425g of the solution from Pass 5 with 713g of water.

Place the hydrom in the vessel containing the mixture.

When it is floating calmly then enter "5" in the Plato field and confirm by pressing "continue".

Session 7: 2.5° Brix/Plato solution

Create a 2.5° Brix solution by mixing 1058g of the solution from Pass 6 with 1058g of water.

Place the hydrom in the vessel containing the mixture.

When it is floating calmly then enter "2.5" in the Plato field and confirm by pressing "continue".

Step 5: Confirm the results

The formula has been calculated and the measured values are displayed in the table. After confirmation the new values are saved in the settings.

The calibration is completed

1

2

3

Nr.

Messung / Lösung

1 2 3 4 5 6 7

Brix soll

°

0,0 20,0 15,0 10,0 7,5 5,0 2,5

Einwagen (g)

Zucker / Wasser vorherige Lsg

0 2100 454 1816 1668 556 1453 727 1619 540 1425 713 1058 1058

Volumen

ml

2100

2100

2100

2100

2100

2100

2100

Dichte Gewicht

bei 20°C g g/ml

1 2100 1,081 2270 1,059 2224 1,038 2180 1,028 2159 1,018 2138 1,008 2117