Do PlainWaterCalibration
The "Plainwater" method is a calibration method that is very fast and uncomplicated. The density of the current liquid is measured and the hydrometer calculates from experience the formula
Prerequisite:
user interface access To make this setting, you must have access to the user interface. If this is not the case, see here how to do it: Access the Userinterface
Equipment needed:
water tank in which the hydrom can float.
place the hydrom in 20 degrees cold water and navigate to "Calibration
Activate the "Plainwater method".
Confirm the calibration with "Start".
The Hydrom measures the current position and compares it with the reference measurement values.
If there is an error message saying that the angle is over >90 degrees, please perform a position sensor calibration. This has only been necessary for a few customers so far and only occurs after the update to 1.5.1. Do Attitude Sensor Calibration
The Hydrom now has a new calibration.
Last updated